Data Logging
REDY CO2 air / water Sensor
Ref : 488 016
Click on OK. The probe calibration is complete.
9.4.2 Second Point Adjustment to 5,82%
In a 30 mL volume beaker or flask :
The probe calibration is complete.
ENGLISH
9.4.2.1
Calibration Solution Preparation
First pour the contents of a pre-dosed ampoule of base solution B
Then add the content of a pre-dosed ampoule of acid solution A
You get a solution equivalent to 5.82% of CO2
9.4.2.2
Second Reference Point Adjustment
✓ Place the probe in the 5.82% solution,
✓ The measured CO2 value increases
✓ Wait for the measurement to stabilize
✓ Click on Calibrate
41