Data Logging
REDY CO2 air / water Sensor
Ref : 488 016
Click on the Measurement tab to check the unit of measurement taken into account :
The selected measurement mode must be Air mode. (For more details, see chapter
8.3)
In principle, a complete calibration will require two reference values:
Click on the Adjustment tab, as specified in the software, put the probe in the air, and wait
for the measurement to stabilize.
Note: You can check the measurement stability, either directly on the sensor screen or by
clicking on the Measurement tab in the software.
ENGLISH
8.4
Complete probe calibration
The first is the level of CO2 in the air, fixed by convention at 0.04%
The second value will be provided by the preparation of a calibration solution
obtained by mixing the contents of an acid solution (pre-dosed bottle A) with that
of a base solution (pre-dosed bottle B).
After stirring, the mixture thus formed gives a known release of CO2 of 5.82%
8.4.1
First Point Adjustment to 0,04%