Data Logging
REDY CO2 air / water Sensor
Ref : 488 016
Then press Display :
Choose
The selected measurement mode must be Air mode. (For more details, see chapter
9.3)
In principle, a complete calibration will require two reference values:
Click on the sensor settings button in the window below
ENGLISH
Air (% unit) or Water (mg / L unit) mode, after pressing the key below:
9.4
Complete Probe Calibration
The first is the level of CO2 in the air, fixed by convention at 0.04%
The second value will be provided by the preparation of a calibration solution
obtained by mixing the contents of an acid solution (pre-dosed bottle A) with that
of a base solution (pre-dosed bottle B).
After stirring, the mixture thus formed gives a known release of CO2 of 5.82%
9.4.1
First Point Adjustment to 0,04%
39