Saving data

manually
Saving data
automatically
Int 1
l Press the <RUN/ENTER> key to continue the three-point calibration.
Std3 appears.
l Enter the current concentration of the third standard solution with
mg/l
< > < >.
l Thoroughly rinse the ISE electrode and, if necessary, the temperature
sensor with deionized water and immerse in the third standard solution.
l Press <RUN/ENTER>. The AutoRead measurement begins. The elec-
trode voltage (mV) appears on the display. If the measured value is stable,
the slope appears in calibration range II.
l Switch to the measuring mode with <M>.
Saving data
Press the <STO> key in the measuring mode (display No. with the number of
the next free memory location). Then, press <RUN/ENTER> and enter the ID
number with < > < >. Terminate the save with <RUN/ENTER>. The instru-
ment changes to the measuring mode.
The portable measuring instrument pH/Ion 340i has an internal data memory.
It can store up to 500 datasets. The save interval (Int 1) determines the chro-
nological interval between automatic save processes.
To set up the save interval, press <STO> while pressing the <RUN/ENTER>
key (display Int 1) and set the interval with < > < > key. Then, press
<RUN/ENTER> and enter the ID number with < > < >. Change to the last
active measuring mode with <RUN/ENTER>. The automatic save is switched
on (display Auto Store).

Outputting the data memory

You can output the data memory with the <RCL> key. By repeatedly press-
ing the <RCL> key, you reach the following functions:
StO dISP
Output stored data on the display
StO SEr
Output stored data on the serial interface
CAL dISP
Output calibration data on the display
CAL SEr
Output calibration data on the serial interface
Start the output with <RUN/ENTER> .
Saving data
11