Transmitting data
Press the <RUN/ENTER> key in the measuring mode.
This manually triggers a data transmission of the current measured value to
the serial interface at any time - independently of the selected intervals.
The interval to the data transmission (Int 2) determines the chronological in-
terval between automatic data transmissions. After the selected interval ex-
pires, the current data record is transmitted to the interface. To set up the
transmission interval, press <RCL> while holding down the <RUN/ENTER>
key (display Int 2). Then, set the interval with < > < >.

Configuring

Note
You can leave the configuration menu at any time with <M>. The parameters
already changed are stored.
Procedure for configuring (factory settings appear in bold typeface):
l Switch off the measuring instrument.
l While pressing the <M> key, press <
l Select the required Baud rate with < > < >.
Selection: 1200, 2400, 4800, 9600 Baud.
l Then press the <RUN/ENTER> key.
l Select the required display during the pH calibration with < > < >.
/pH
Selection: Buffer nominal value (/pH) or current electrode voltage (mV).
l Then press the <RUN/ENTER> key.
l Set the required calibration interval with < > < >.
Selection: 1... 7... 999 d.
l Then press the <RUN/ENTER> key.
l Select the date and time step-by-step with < > < >. In between, press
the <RUN/ENTER> key each time.
l When the last <RUN/ENTER> key has been pressed, the instrument
changes to the last active measuring mode.
>.