Saving Data; Outputting The Data Memory; Erasing The Data Storage - wtw Oxi 1970i Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Oxi 1970i

Saving data

manually
Saving data
automatically Int 1
Saving data
The measuring instrument has an internal data memory. It can store up to
800 datasets.
Press the <STO> key in the measuring mode (display No. with the number of
the next free memory location). Then, press <RUN/ENTER> and enter the ID
number with <▲> <▼>. Finish the save with <RUN/ENTER>. The instrument
switches to the measuring mode.
The save interval (Int 1) determines the chronological interval between
automatic save processes.
To set up the save interval, press <STO> while pressing the <RUN/ENTER>
key (display Int 1) and set the interval with <▲> <▼>. Then press <RUN/
ENTER>. Press <RUN/ENTER> once more and enter the ID number with
<▲> <▼>. Switch to the last active measuring mode with <RUN/ENTER>.
The automatic save is switched on (display Auto Store).

Outputting the data memory

You can output the data memory with the <RCL> key. By repeatedly pressing
the <RCL> key, you reach the following functions:
StO dISP
Output stored data on the display
StO SEr
Output stored data on the serial interface
CAL dISP
Output calibration data on the display
CAL SEr
Output calibration data on the serial interface
Start the output with <RUN/ENTER>.

Erasing the data storage

Switch off the measuring instrument. Press the <STO> key and keep it
depressed. Press the <ON/OFF> key. Confirm the erasing procedure with
<RUN/ENTER> or cancel the erasing procedure by pressing any other key.
Saving data
23

Publicité

Table des Matières
loading

Table des Matières