Power measurements require simultaneous measurement capability (refer to
If a FETCh query is sent before the measurement is finished, the response will be delayed until the
measurement trigger occurs and the acquisition completes. You can test the MEAS_active bit in the
operation status register to know when the measurement trigger system has returned to the idle
state.
Front Panel Menu Reference
Not available
If a bit value of 32 is returned in the query, the MEAS_active bit is true, and the measurement is NOT
complete. When the MEAS_active bit is false, you can retrieve the measurement. Refer to Status
Tutorial in the Operating and Service Guide document for more information.
Keysight N6700C Series User's Guide
SCPI Command
To query the MEAS_active bit (bit 5):
STAT:OPER:COND? (@1)