4 Operating the Power System
Front Panel Menu Reference
Select Transient\Control.
Navigate to Initiate. Then press Select.
It takes a few milliseconds for the instrument to be ready to receive a trigger signal after receiving the
INITiate:TRANsient command.
If a trigger occurs before the trigger system is ready for it, the trigger will be ignored. You can test the
WTG_tran bit in the operation status register to know when the instrument is ready to receive a
trigger.
Front Panel Menu Reference
Not Available
If a bit value of 16 is returned in the query, the WTG_tran bit is true, and the instrument is ready to
receive the trigger signal. Refer to Status Tutorial in the Operating and Service Guide document for
more information.
Unless INITiate:CONTinuous:TRANsient is programmed, it will be necessary to initiate
the transient trigger system each time a triggered action is desired
Trigger the output step
The trigger system is waiting for a trigger signal in the initiated state. You can immediately trigger the
output as follows:
Front Panel Menu Reference
Select Transient\Control.
Navigate to Trigger. This generates an immediate trigger
signal regardless of the trigger source setting. Then
press Select.
As previously discussed, a trigger can also be generated by another output channel or by a trigger
signal applied to an input pin on the digital port connector. If any of these systems are configured as
the trigger source, the instrument will wait indefinitely for the trigger signal. If the trigger does not
occur, you must manually return the trigger system to the idle state.
The following commands return the trigger system to the idle state:
Front Panel Menu Reference
Select Transient\Control.
Navigate to Abort. Then press Select.
104
SCPI Command
To initiate the transient trigger system:
INIT:TRAN (@1)
SCPI Command
To query the WTG_tran bit (bit 4):
STAT:OPER:COND? (@1)
SCPI Command
To generate a transient trigger:
TRIG:TRAN (@1)
Alternatively, you can also program a *TRG or an IEEE-
488 <get> command.
SCPI Command
ABOR:TRAN (@1)
Keysight N6700C Series User's Guide