Télécharger Imprimer la page

Endress+Hauser Liquicap M FMI51 Mode D'emploi page 16

Masquer les pouces Voir aussi pour Liquicap M FMI51:

Publicité

Les langues disponibles

Les langues disponibles

BVS 05 ATEX E 103 X / XA327F-F
Safety instructions:
Zone 0
Zone 0/1 Application:
16
• Only operate devices in potentially explosive gas-air and/or dust-air mixtures under atmospheric
conditions: –20 °C ≤ T ≤ +60 °C
0.8 bar ≤ p ≤ 1.1 bar
• If no potentially explosive mixtures are present, or if additional protective measures have been taken,
according to EN 1127-1, the transmitters may be operated under other atmospheric conditions in
accordance with the manufacturer's specifications.
• Install the device to exclude any mechanical damage or friction during the application. Pay particular
attention to flow conditions and tank fittings.
• Mechanically fix probes which are more than 3 m.
Electronic
Type of
insert
protection
FEI50H
II 1/2 G
Ex ia IIC/IIB
FEI57C
II 1/2 G
Ex ia IIC/IIB
FEI55
II 1/2 G
Ex ia IIC
FEI57S
II 1/2 G
Ex ia IIC
FEI58
II 1/2 G
Ex ia IIC
* for F16 housing:
–40 °C...+55 °C, +60 °C and/or +70 °C
Electrical data Ambient temperature
electronic
Ui ≤ 30 V
–50 °C ≤ Ta ≤ +60 °C *
Ii ≤ 120 mA
–50 °C ≤ Ta ≤ +70 °C *
Pi ≤ 1 W
Ci ≤ 2.4 nF
Li ≤ 0
Ui ≤ 19.2 V
–50 °C ≤ Ta ≤ +60 °C *
Ii ≤ 108 mA
–50 °C ≤ Ta ≤ +70 °C *
Pi ≤ 1 W
Ci ≤ 2.4 nF
Li ≤ 0
Ui ≤ 36 V
–50 °C ≤ Ta ≤ +55 °C *
Ii ≤ 100 mA
–50 °C ≤ Ta ≤ +70 °C *
Pi ≤ 1 W
Ci ≤ 2.4 nF
Li ≤ 0
Ui ≤ 16.1 V
–50 °C ≤ Ta ≤ +55 °C *
Ii ≤ 100 mA
–50 °C ≤ Ta ≤ +70 °C *
Pi ≤ 1 W
Ci ≤ 2.4 nF
Li ≤ 0
Ui ≤ 18 V
–50 °C ≤ Ta ≤ +60 °C *
Ii ≤ 52 mA
Pi ≤ 170 mW
Ci ≤ 0
Li ≤ 0
Liquicap M
Temperature
Process
class
temperature
T6
T3...T5
see the
temperature
diagram on
T6
the next page
T3...T5
T6
T3... T5
T6
see the
temperature
T3...T5
diagram on
the next page
T3...T6
Endress+Hauser

Publicité

loading

Ce manuel est également adapté pour:

Liquicap m fmi52Liquicap m fti51Liquicap m fti52