Operation
Retraction function
By a short return stroke of the pump, the retraction function avoids dripping in dosing
mode.
To activate the retraction function, follow these steps:
Turn off the device.
Press the [Cal] button.
While holding down the [Cal] button, turn the device on again.
The display shows the currently set pump head.
Release the [Cal] button and press one of the arrow buttons until
(retraction function active) or
Confirm with [Select].
Fast filling/emptying
With the [max] function, the pump can be temporarily set to maximum power for fast
filling/emptying of the connected tube(s), independent of all other setting values:
Turn on the device as described in section „Switching the device on/off" auf Seite
64
Make sure that the rotation movement is stopped.
Use the [reversing] button to select the desired direction of rotation of the pump
head (clockwise/counter-clockwise, observe the direction arrows on the device).
Press and hold the [max] button.
The pump runs at maximum speed to fill/drain the connected tube(s).
The pump stops as soon as you release the button.
[r off]
(without retraction) is displayed.
70
[r on]