Follow the instructions below to keep the unit in good condition.
Store the unit in the case when not in use.
Do not fold the wrist sleeve too hard.
The Velcro strip can damage the inside surface of the sleeve.
Clean the unit and sleeve with a soft, dry cloth.
Do not use abrasive or solvent-based detergents.
IMPORTANT:
Do not immerse the unit or its components in water or
any other liquid. Do not expose the unit to moisture,
extreme temperatures or direct sunlight.
Store the unit and its components in a safe place.
Do not expose the unit to knocks and blows.
Remove the batteries if the unit is not going to be used
for 3 months, or more. Replace all the batteries at the
same time.
The troubleshooting chart will help you identify the causes of common problems and how they can be
rectified. Contact your dealer if you cannot rectify the problem by using the troubleshooting chart.
PROBLEM/ERROR CODE
Nothing is displayed when you
press the power switch, or the
battery indicator
flashes
E1: The air pressure does not
rise correctly
E3: The inflation pressure is
too high
E2, E4: Movement during the
measuring
The battery indicator is
displayed
ENGLISH
MAINTENANCE
TROUBLESHOOTING
POSSIBLE CAUSE
No batteries
The batteries are
discharged.
The batteries have the
wrong polarity
The blood pressure
sleeve is leaking
The arm or body moved
during the measurement
The batteries are
discharged.
40
ACTION
Insert the batteries
Insert new batteries
Insert the batteries with the correct
polarity
Replace the sleeve
Contact your dealer
Sit still in the correct posture and
measure again
Insert new batteries