pH-value with Tablet

6.5 – 8.4
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Add one PHENOL RED PHOTOMETER tablet straight from the foil to
the 10 ml water sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
Place the vial in the sample chamber making sure that the
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display as pH-value.
Tolerance: ± 0.1 pH