Functional description

GB
Zero
Test
METHOD
RESULT
Zero
Test
Zero
Test
!
!
!
Zero
Test
MD100_1 02/2018
After zero calibration is completed, remove the vial from the sample
chamber. The characteristic coloration appears after the addition of
the reagents.
Replace the cap on the vial and place in the sample chamber making
sure that the
marks are aligned.
Press the [ZERO/TEST] key.
(For Countdown /reaction period see page 33)
The "Method" symbol flashes for approx. 3 seconds.
The result appears in the display.
The result is saved automatically.
Repeating the test:
Press the [ZERO/TEST] key again.
Repeating the zero:
Press the
[ZERO/TEST]
key for 2 seconds.

Display backlight

Press the [ ! ] key to turn the display backlight on or off. The backlight
is switched off automatically during the measurement.

Recall of stored data

If the instrument is switched on, press the [!] key for more than 4 sec-
onds, then release the [!] key to access the recall menu.
Countdown / reaction period
If a reaction period is included in a method a countdown function
can be used:
Press the [!] key and hold.
Press the [ZERO/TEST] key.
Release the [!] key; the countdown starts.
After the countdown is finished the measurement starts automatically.
It is possible to interrupt the countdown by pressing the [ZERO/TEST]
key. Measurement starts immediately.
Caution:
An incomplete reaction period can lead to incorrect test results.
33