Cleaning the circuit

1
It is recommended to clean the liquid circuit daily.
Cleaning is necessary to reduce crystalline deposits and the number of bacteria in the liquid circuit of
the unit; the following procedure achieves a 5 log reduction of target pathogens (using washing solutions
for cleaning the circuit as listed at the beginning of this manual):
1
Remove the instrument,
2
Pre-Cleaning the circuit during 20 seconds with distilled (or demineralized) water,
3
Cleaning the circuit during 20 seconds with an appropriate washing solution for washing the circuit
4
Allow the washing liquid to act for 5 minutes,
5
Rinse the circuit during 20 seconds with sterile water.
50
All manuals and user guides at all-guides.com
2
20 s
H
O
2
(distilled water)
3
20 s
disinfecting
liquid