7.6
Inspection, maintenance and checks
Inspect the product after each cleaning and disinfecting cycle to be
sure it is clean, functional and undamaged.
Set aside the product if it is damaged.
9.
Troubleshooting list
Have defective products repaired by Aesculap Technical Service, see Technical Service.
Issue
Charger not working
Defective indicator
Battery does not charge
Detection
"Power On" indicator not illumi-
nated
During the switch-on check, not all
indicator elements are illuminated
("Power On" indicator and indicator
fields for charging bays )
Battery inserted, charging progress
indicator not illuminated
Battery inserted, "Action request"
indicator illuminated
8.
Maintenance
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
Cause
Mains power cord not plugged in
Power cord faulty
Fuses blown
Charger defect
Charger defect
Charger contacts dirty
Battery defective
Charger defect
Excessive battery temperature
detected during charging
Battery defective
Charger defect
Remedy
Plug mains power cord into the
device connector 7 on the charger
and into the socket of the mains
power supply.
Replace power cord.
Replace fuses, see Replacing fuses.
Replace product.
Replace product.
Clean the contacts, see Validated
reprocessing procedure.
Repeat the charging process in
another charging bay. If the prob-
lem occurs again, replace product.
The charging process does not start
up either with a different battery.
Replace product.
Remove the battery from the
charging bay, let it cool down and
restart charging. If the problem
occurs again, replace battery.
Repeat the charging process in
another charging bay. If the prob-
lem occurs again, replace battery.
The charging process does not start
up either with a different battery.
Replace product.