FUSE HOLDER AND VOLTAGE CHANGER
ASSEMBLY
If a change in line voltage operation is desired, proceed
as follows:
Unplug the power cord from the controller rear
panel socket.
On the power entry module (rear panel), check back
door for voltage selector set.
Using a small screwdriver, pull out the voltage
selector and fuses by levering in position A.
Replace the fuse.
Use only T-type fuses of the following
characteristics:
- 100/120 Vac 1 A
- 220/240 Vac 1 A
FUSE HOLDER
AND VOLTAGE
SELECTOR
Rear panel
Select the operating voltage, then firmly insert the
voltage selector and fuses in place.
Voltage Changer
Check voltage selector window for correct set and
connect the power cord.

Input/Output Interconnections

All input/output signals are accessible at J001
accessory connector.
With the provided P001 mating connector make the
connections with AWG 24 (0.24 mm 2 ) - or smaller wire
- to the pins indicated in following figure, to obtain the
desired capability.

Input Signals

Pin 3-7 Remote high voltage ON/OFF or high voltage
Pin 3-7 must be shorted to allow the high voltage
output if no remote interlock contact is connected.

Output Signals

The front panel REC.OUT connector is a duplicate of
this signal on a BNC connector.
PIN
2-6
A
4-8
5-9
21
TECHNICAL INFORMATION
interlock; requires a permanent closed pin
contact.
NOTE
DESCRIPTION
Current recorder output (pin 2 signal out, pin
6 ground). The output voltage
+10 Vdc is proportional to the pump current 0
to 1 mA (e.g. 10 Vdc correspond to 1 mA).
The resolution is 10 mV minimum.
Voltage recorder output (pin 4 signal out, pin
8 ground). The output voltage 0 to 3 V is
proportional to the controller high voltage
output 0 to 3 kV (e.g. 1 V corresponds to 1
kV in linear mode).
High voltage indication contact. This pure
contact (which carries 1 A at 250 Vac and 0.2
A at 30 Vdc) is open when the high voltage is
off and closes when the high voltage in on.