Comply with the statutory provisions and the current state-of-the-
art when cleaning and disinfecting the device. Use the detergent
and disinfectant recommended by the manufacturer to clean and
disinfect the device. Clean and disinfect your device and access-
ries in line with the specifications in Chapter
maintaining the device

OPERATION, MAINTENANCE AND CARE

Metrological control (MTK): the device is calibrated by the manu-
facturer for a period of two years. The metrological control must take
place in accordance with §14 MPBetreibV (Medical Device Opera-
tor Regulation) at least every two years and after repairs. The inspec-
tion is subject to a charge and may be performed by the manufactu-
rer or an authorised maintenance service centre, in accordance with
the Medical Device Operator Regulation.
Every safety check must be completed in full and consists of the
following components: visual inspection, functional check, testing of
the monitoring, safety, display and messaging systems, measurement
of the key safety values, electrical test.
Only use the device in accordance with its intended use defined
in the instructions for use. The warranty expires if the device is used
incorrectly.
The compression or other mechanical restriction of the cross-
section of connecting tubes must be avoided.
The device may not be used together with HF surgery, x-rays and
MRI.
and
7.4. Cleaning and maintaining the cuff.