Adjustment Ranges - Knick SensoTrans A 20220P0 Serie Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Zero Point:
Use DIP4, DIP5, and DIP6 to set the number to the left of the decimal
point for the zero point. Use the "Zero" rotary switches (Z) to set the
numbers to the right of the decimal point (x.00 ... x.99).
Range
DIP S1
4 5 6
–(3+Z/100)
–(2+Z/100)
x
–(1+Z/100)
x
–(0+Z/100)
x x
0+Z/100
x
1+Z/100
x
x
2+Z/100
x x
3+Z/100
x x x
Within these ranges the exact value is specified by
the rotary switches ("Z").
Output Signals:
Adjust the output signal using the switches DIP7, DIP8.
Note
Cover the switches with the included self-adhesive polyimide tape
after completion of confi guration.
-3.99
-2.99
-1.99

Adjustment Ranges

-0.99
0
0.99
1.99
2.99
3.99
[mV/V]
39

Publicité

Table des Matières
loading

Table des Matières