Transport Control This sets the mattress into Transport mode where the
Deflating the
Mattress
mattress is sealed and the support surface is equally
pressurised; the pump and/or tubeset can then be
removed. In this mode the mattress will support the
patient for up to 12 hours.
To set Transport mode:
1. Turn the Transport control knob clockwise to
Transport.
2. Put the pump into Standby and disconnect the
tubeset.
If the Transport control is set to Transport with the
tubeset connected and the pump switched on, the
pump will indicate a Low Pressure fault alarm.
To resume normal operation:
1. Re-connect the pump and tubeset to the mattress.
2. Turn the Transport control knob counterclockwise
to Normal.
3. Put the pump into the Run mode.
4. Check the system is operating normally.
To deflate and store the mattress, do the following:
1. Put the pump into Standby, and disconnect the
pump from the mains power supply.
2. Remove the tubeset from the pump and mattress.
3. Activate the CPR control to deflate the mattress.
4. Make sure the Transport control is set to Normal.
5. Fold the mattress over in the middle to assist in air
evacuation; if necessary, gently press on the base
cover to increase the air loss.
6. Roll up the mattress, starting at the foot end.
Make sure the mattress is dry before rolling it up.