5.0 MAINTENANCE

5.1

Microscopy environment

This microscope is recommended to be used in a clean, dry and shock free environment with a tem-
perature of 0-40°C and a maximum relative humidity of 85 % (non condensing). Use a dehumidifier if
needed.
5.2
To think about when and after using the microscope
The microscope should always be kept vertically when moving it and be careful so that no moving
parts, such as the eyepieces, fall out.
Never mishandle or impose unnecessary force on the microscope.
Never attempt to service the microscope yourself.
After use, turn off the light immediately, cover the microscope with the included dust-cover, and keep
it in a dry and clean place.
5.3
Electrical safety precautions
Users should observe all safety regulations of the region. The equipment has acquired the CE safety
label. However, users do have full responsibility to use this equipment safely.
5.4
Cleaning the optics
If the optical parts need to be cleaned try first to: use compressed air.
If that is not sufficient: use a soft lint-free piece of cloth with water and a mild detergent.
And as a final option: use the piece of cloth moistened with a 3:7 mixture of ethanol and ether.
Note: ethanol and ether are highly flammable liquids. Do not use them near a heat source, near
sparks or near electric equipment. Use these chemicals in a well ventilated room.
Remember to never wipe the surface of any optical items with your hands. Fingerprints can damage
the optics.
Do not disassemble objectives or eyepieces in attempt to clean them.