Available Spare Parts - FLAEM AirMate Mode D'emploi

Masquer les pouces Voir aussi pour AirMate:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 10

AVAILABLE SPARE PARTS

Description
- RF8 Nebuliser +Mouthpiece
- Mask
- Connection tube (1 m)
- Air filter (2 pcs)
TROUBLE-SHOOTING
Switch off the device before any procedure and unplug the power cable from the
socket.
PROBLEM
The device does not work
The device does not
nebulize or
nebulizes insufficiently.
The apparatus is noisier
than usual
If after verifying the above mentioned conditions the device should not operate properly,
contact your trusted retailer or the nearest authorized FLAEM service center.
CAUSE
The power cable has not
been correctly inserted into
the power socket
The medication has not been
inserted in the nebulizer
The nebulizer has not been
properly fitted.
The nebulizer nozzle is
clogged
The air pipe is not correctly
connected to the device
The air pipe is bent, damaged
or twisted
The air filter is dirty
The Filter is not correctly
inserted in its housing
5
Code
ACO597P
ACO512P
ACO35
ACO439P
SOLUTION
Correctly insert the power
cable in sockets.
Pour the right amount of
medication in the nebulizer
Disassemble the nebulizer and
reassemble it correctly as
shown in the connection
diagram on the cover.
Disassemble the nebulizer,
remove the nozzle and
perform cleaning operations.
Failure to remove medication
deposits from the nebulizer
affects its efficiency and
operation.
Strictly comply with the
instructions
contained in the "CLEANING,
SANITISATION, DISINFECTION"
chapter
Ensure the device's accessories
are properly connected to
the air intake (see connection
diagram on the cover).
Unwind the pipe and ensure
it is
not crushed or punctured. If
necessary replace it.
Replace the filter
Fully and properly insert the
Filter in its housing

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sc36p00

Table des Matières