4.3 Environmental conditions and location of the equipment
Temperature
Relative humidity
Altitude
Pollution level
Other environmental
conditions
Equipment and tank location
4.4 Consumable storage requirements
Use original consumables. The use of consumables from another source may result in damage to the equipment and
the loss of the warranty. Check the consumable storage indications in the documentation delivered with the
equipment.
4.5 Quality of the final water and equipment performance
Conductivity
(compensated at 25ºC)
a
Production
Maximum recommended
production
Silica removal
Reverse
osmosis
performance
a
The production values depend on the model. 15% fluctuation. The production values are based on standard
operating conditions [25ºC / pH 5 - 7/4 bar / 250 ppm NaCl] and can vary according to the quality of the feed water and
the typical variations of the reverse osmosis membranes
4.6 Dimensions and weight
Weight in operation of the
equipment
Equipment dimensions
(Height/Width/Depth) [cm]
Working: 5°C – 35°C. Ideally, the working temperature lies in a 20-25°C range.
Storage: 5°C – 35°C.
No condensation: < 80% for temperatures up to 31°C (linear decrease to 50% at
40°C).
<3000 m
1 (No pollution or only dry, nonconductive pollution).
Avoid direct sunlight.
Avoid aggressive environments (chemical agents, dust, dirt).
The equipment is not intended for use in explosive atmospheres.
Indoors.
On a flat surface. Ensure the flat surface is able to support the weight of the
equipment or tank (See Section 4.6 Dimensions and weight)
Always keep the equipment upright. In case of transportation or in case of moving
the equipment, disconnect the cartridges to avoid damaging the connectors during
transportation. To move the equipment, hold it by the base.
The equipment needs a drainage point for the rejection water. The drainage point
must be located at a maximum distance of 3 m.
The feed water connection point should be located at a maximum distance of 3 m.
The connection point to the electrical network should be located at a maximum
distance of 2 m.
Locate the tank considering that it is the point of use of Type II water. The tank must
be preferably opaque and with a lid. The tank should not receive direct sunlight. It is
recommended to have a safety overflow line directed to a drainage point and to have
enough space above it in order to be able to easily clean the tank.
< 1 µS/cm
2.5 L/h
25 L/day
>99,9%
90– 98%
9 kg
45 x 25 x 40
8/31