Using the STORE function
STORE/RECALL button:
Measurement results can be stored in the tester memory for subsequent recall
to the display.
Press the STORE/RECALL button once to store the voltage value during a
voltage test.
Press the STORE/RECALL button once to store the PI/ DAR, insulation or
resistance value when the individual tests are completed.
The data storage location number the "MEM" symbol will flash twice when the
data is stored.
Note. The memory is divided into five segments, with each segment able to
store up to a maximum of 100 pieces of data. The method to store the data is
first in / first out when the memory is full.
The stored values of the individual test
segments
Voltage
Items
1
Voltage
2
3
NOTE:
Press the STORE/RECALL button ≥5 sec to clear the stored data.
The display will flash the "MEM" symbol and " cLr " twice.
Using the RECALL function
Press the STORE/RECALL button ≥2 sec to enter/exit the RECALL mode.
If the memory is empty, the tester will display the "nOnE" symbol.
Search the stored value under RECALL mode
MEM
16
Insulation
Earth-Bond
resistance
resistance
50~1000V
Resistance
Resistance
Leakage
current
Test voltage
Press
M
Press
1
COMP
DAR
DAR value
R30-sec
R1-min
MEM
M
2
MEM
M
3
PI
PI value
R1-min
R10-min