Commissioning, Operation; Mechanical Mounting; Safety Inspection - WIKA IL-10 Mode D'emploi

Transmetteur de pression immergeable à sécurité intrinsèque
Masquer les pouces Voir aussi pour IL-10:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5. Commissioning, operation

5. Commissioning, operation

5.1 Mechanical mounting

EN

5.1.1 Safety inspection

Only use the instrument if it is in perfect condition with respect to safety.
5.1.2 Requirements for the mounting point
The existence of strong electromagnetic fields in a frequency range of < 2.7 GHz can result in increased measuring
errors up to 1 % of span. Do not install the instruments in the vicinity of strong electromagnetic sources of interference
(e.g. transmitting devices, radio equipment), or use sheath current filters where applicable.
5.1.3 Mounting the instrument
Maximum tensile force of the FEP cable:
350 N without strain relief
500 N with strain relief
Maximum tensile force of the PUR cable:
350 N without strain relief
1,000 N with strain relief
The protection cap protects the internal diaphragm from
damage during transport and during the lowering of the
probe. Remove the protection cap if the medium is viscous
or contaminated.
Protect the diaphragm from contact with abrasive media
and against any impacts.
16
Prior to commissioning, the instrument must be subjected to a visual inspection.
Leaking fluid is indicative of damage.
With a damaged diaphragm, explosion protection cannot be guaranteed.
WIKA operating instructions submersible pressure transmitter model IL-10
Protection cap
Diaphragm

Publicité

Table des Matières
loading

Table des Matières