TROUBLESHOOTING (CONTINUED)
ERROR LIST
ERROR
NO.
Err 1
Err 2
Err 3
Err 5
Err 6
Err 7
DETAILED VIEWS
DEFINITION
Mathematical mistake; division by zero.
Cannot calculate calibration factor
Weight loaded too low. Cannot calibrate
Unstable weight. Cannot calibrate
Calibration disabled
High saturation state (high A/D)
Low saturation state (low A/D)
CHECK/REPLACE
Will show when trying to calibrate with no
calibration weight
While in calibration procedure add more weight in
load mode
Check the load cells mechanical surroundings and
see nothing touches them and the cables are well
connected
Toggle the calibration switch to enable
The load cell is not connected correctly. Check
the connection, cables, and mechanical
connection. If the problem persists, replace the
set of load cells
See Err 6