3. Error Codes

Important! By displaying the following error codes the instrument support the
operation of the instrument.
Error
Meaning
oPEn
no probe or wrong probe is connected
"too low" – underflow of the measuring range
„too high" – exceeding of the measuring range
temperature of the cold junction above the measuring range
temperature of the cold junction above the measuring range
E15
low battery – (battery is com-pletely exhausted)
E19
low battery – (battery is com-pletely exhausted)
E16
time out / auto-off-function
E1
time out / auto-off-function
E1e
time out / auto-off-function
E23
EE-prom destroyed or EE-prom index is wrong
E25
EE-prom destroyed or EE-prom index is wrong
E31
temperature of cold junction out of range
LibA
Internal buffer battery (CR2032) empty
(for changing the battery the housing must be unscrewed)
rEr1
maximum statistical value cannot be calculated
rEr2
maximum statistical value cannot be calculated
Notice:
Defective or empty batteries can lead to measurement errors or error messages
produce. In the event of error messages, remove the battery and wait approx.
3 minutes before inserting a new battery.
34

4. Technical data

P700/P700B (1-channel) / P705 (2-channel):
Inputs 1/2
Pt100, thermocouple: Typ K, J, L, N, R, S, T
Measuring range
Pt100
-200...+850 °C
Thermocouple
according DIN (-200...1760°C)
Accuracy
Pt100
±0,1 °C from -100 °C...+200 °C 0,1%
Thermocouple
R, S
±1,0 °C +0,1%
K, J, L, N, T
±0.2 °C from 0 °C...+200 °C
±0.5 °C to 1000 °C
±1.0°C remaining range
Resolution
0,1 °C
Ex-mark
-
Battery life
appr. 20 hours
Connectors
DIN 8-pole
Allow. operating temp. 0 °C ... +40 °C
Display
2-line LCD
Housing
plastic
Dimensions
200 x 93 x 44 mm (LxWxH)
Weight
350 g
Power supply
9 V battery
P710 (1-channel) / P715 (2-channel)
Inputs 1/2
thermocouple: Typ K, J, L, N, R, S, T
Measuring range
Pt100
-200...+850 °C
Thermocouple
according DIN(-200...1760°C)
Accuracy
Pt100
-
Thermocouple
R, S
-
K, J, L, N, T
±0.2 °C from 0 °C...+200 °C
±0.5 °C to 1000 °C
±1.0°C remaining range
Resolution
0.1°C
Ex-mark
-
Battery life
appr. 20 hours
Connectors
DIN 8-polig
Allowable operating temp. 0 °C ... +40 °C
Display
2-line LCD
Housing
plastic
Dimensions
200 x 93 x 44 mm (LxWxH)
Weight
350 g
Power supply
9 V battery
remaining range
35