3. Programming and use
Example: Correction of the accuracy
Quantity dispensed: 1,6 l / Quantity displayed: 1,52 l
Correction factor k = 1,6/1,52 = 1,0526
Calculation of the new correction factor: 0,9950 x 1,0526 = 1,0473
1.
3.
5.
7.
9.
11.
Measuring unit
„Total" resettable
„Total" not resettable
Press 3 times „Total" and 3 times
„Reset" to access the
programmation mode. Measuring
unit "L" (liter) is flashing and can
be changed in L, GAL, QT or PT
when pressing "Reset". Press
"Total" to confirm the new
measuring unit.
The second figure can now be
chanced.
To access the next figure, press
"Total".
The fourth figure can now be
chanced.
To access the next figure, press
"Total".
Press both buttons simultaneously
to save the programmation.
The meter then goes to the
sleeping mode.
After the sleeping mode,
please press „Total" to
show the „normal" display
mode.
2.
Press „Total" one time to
switch on the meter.
4.
Press "Total" one time to change the
k-factor.
The figure to be changed is flashing
and can be modified with "Reset".
To change the next figure, press
"Total".
6
8.
10.
12.
The third figure can now be
chanced.
To access the next figure, press
"Total".
The fifth figure can
now be changed.
Sleeping mode
s
Press „Reset" during 3 seconds
and the register displays the
check sum.
7