Servicing

8.1. Servicing Schedule

Component
Pump
Filter (Accessory)
Diaphragm and valve
plates/sealings (valve
plates)
Tab. 9
8.2. Cleaning
When cleaning, make sure that no liquids enter the inside of
the housing.
8.2.1. Flushing Pump
 Before switching off the pump, flush it with air (if neccesary for
safety reasons: with an inert gas) for about five minutes under
atmospheric conditions (ambient pressure).

8.2.2. Cleaning Pump

 As far as possible, clean the parts with a dry cloth.
 Only use solvents for cleaning if the head materials cannot be
attacked (check the resistance of the material!).
 If compressed air is available, blow out the components.
8.3. Changing Diaphragm and Valves
8.3.1. N 86 K_.18
Pump is switched off and mains plug is removed from the
socket
Pump is clean and free of hazardous materials
Tubes removed from pump's pneumatic inlet and outlet
Spare part/tool
Service Set (according to chapter 10)
Philips-head screwdriver No. 1
Small screwdriver, blade width 0.5 mm
Pencil
Tab. 10
 Always replace diaphragm and valve plates together to main-
tain the pump performance.
Servicing interval
Regular inspection for external damage
or leaks
Replace if it is dirty
Replace at the latest, when pump
output decreases
Servicing
DE
EN
FR
ES
IT
17