KNF Laboport N86 KN.18 Mode D'emploi page 48

Masquer les pouces Voir aussi pour Laboport N86 KN.18:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Servicing
Conditions
Spare parts/Tools
Information on procedure
Fig. 7: Removing pump head
20
Translation of original Operating Instructions, English, KNF 121435-121438 04/18
8.3.2. N 811 K_.18
Pump is switched off and mains plug is removed from the
socket
Pump is clean and free of hazardous materials
Tubes removed from pump's pneumatic inlet and outlet
Spare part/tool
Service Set (according to chapter 10)
Philips-head screwdriver No. 1
Small screwdriver, blade width 0.5 mm
Pencil
Tab. 11
 Always replace diaphragm and valve plates/sealings together
to maintain the pump performance.
Health hazard due to dangerous substances
in the pump!
Depending on the substance transferred, caustic
burns or poisoning are possible.
WARNING
 Wear protective clothing if necessary, e.g.
protective gloves.
 Flush pump before replacing the diaphragm and
valve plates/sealings (see chapter 8.2.1).
Removing pump head
1. Mark the position of head plate (Fig. 7/3), intermediate plate
(2), cover (5) and cover plate (6) by a drawing line with a felt-
tip pen. This helps to avoid incorrect assembly later.
2. Undo the 4 screws (4) in the head plate (3) and lift the head
plate together with intermediate plate (2) off the pump hous-
ing.
Mini Laboratory pumps N 86.18, N 811.18