IMPORTANT INSTRUCTIONS: USER OPERATING MANUAL
Warming Temperature Adjustments: Activate warming air mode by pressing any temperature
preset button. Once active, warming air temperature can be further adjusted by 1°C increments
from 43°C to 32°C using the +/‐ buttons. Audio announcement indicates current temperature
setpoint.
Auto‐shutoff Timer Adjustment: Activation of any air mode defaults the auto‐shutoff timer to 4
hours. Once active, auto‐shutoff time can be further adjusted by 30‐minute increments using the
+/‐ buttons. 4 hours auto‐shutoff is the maximum allowable time setting. Audio announcement
indicates current auto‐shutoff timer setting.
Fan Speed Adjustments: Activation of any mode defaults to 100% maximum fan speed. For
smaller animals with smaller warming bags, the fan speed may be reduced to lower the noise of
the unit. The unit has 20 fan speed settings and will stop beeping when upper or lower fan speed
limits are reached using the +/‐ buttons.
Programmable One‐Touch Memory Recall: Frequently used settings can be recalled using the
one‐touch memory buttons. Simply turn the unit on and set to preferred temperature, fan speed
and auto‐shutoff settings and then press and hold any memory key for 3 seconds to save. When
the unit is in standby, press and quickly release the same memory key to activate the unit and
recall the saved settings.
HEPA Filter Change Reset: The unit contains a permanent service clock of operating hours. When
400‐hour intervals are reached, a "CHANGE FILTER" audio alert will be heard. Reset the "CHANGE
FILTER" audio alert after the HEPA filter has been changed (See Maintenance section) by
simultaneously pressing the setup button (gear icon) and OFF button the remote control.
Revision 1.0
MODEL SKU: VET‐800NA
Status Display Ring Indicator
Solid Red = Operating at set air temperature
Spinning Red = Adjusting to reach set air temperature
Solid Blue = Cool mode
P a g e | 8