DISPLAY
WHAT IT MEANS
The setting alarm is ringing
as a reminder to do the
blood glucose test.
The meter is in the
PC communication
mode.
OTHER PROBLEMS THAT MAY OCCUR
The test strip has not been
inserted into the meter properly.
Defective test strip.
Insufficient blood sample.
Test strip remains in the
test port for more than 3
minutes prior to testing.
LCD display on the meter is blank
when trying to perform a test.
ACTION
Press M or S button to shut
off or the alarm will turn
off automatically after 30
sec.
Follow
the
provided in the software to
download the test
from the meter.
ACTION
Review the instructions and re-
insert the test strip correctly
(white side face up).
Replace with a new test strip.
Repeat the test with new test strip.
Meter will automatically turn-off. Re-insert
the test strip into the test port.
Contact Customer Service for help.