Maintenance; Troubleshooting - SCHOTT VisiLEDs MC Serie Conseils D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

The VisiLED MC series is maintenance-free.
There is no provision necessary for disinfecting the MC 750 or MC 1500
controller as well as the VisiLED illumination units when using them in the
medical field.
To clean the outside of the controller or the illumination components please
use a soft dry cloth or commercially available plastic cleaning cloths.
The use of cleaning agents, alcohol or any other chemicals is not permitted.
Fault
Possible causes and remedial action
1. No light,
operation display (19)
not lit up
2. No light,
operation display (19)
lit up
3. No light,
red LED display (8)
permanently lit up
4. Light available,
red LED display (8)
blinking
Is electronic power supply correctly connected
to LED controller and power cord? (16)
Is correct supply voltage available?
Is controller switched on? (17)
Is LED illumination connected correctly? (11)
Is brightness setting too low? If necessary,
increase by turning knob (18)
LED unit heated up excessively.
Reduce brightness (18), until the red LED goes
out. Also see Section 7.6
Warning that the LED illumination is
approaching the permitted temperature limit.
Also see Section 7.6

9. Maintenance

10. Troubleshooting

MC 750
VisiLEDs MC Serie | 53
GB

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Visileds mc1500Visileds mc750

Table des Matières