Guidance and manufacture's declaration – electromagnetic immunity –
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture's declaration – electromagnetic immunity
The CMS70A is intended for use in the electromagnetic environment specified below. The
customer or the user of CMS70A should assure that it is used in such an environment.
Immunity
IEC 60601
test
test level
Conducted
3Vrms
RF
150 kHz to 80
IEC
MHz
61000-4-6
outside
bands
Radiated
3 V/m
RF
80 MHz to 2.5
IEC
GHz
61000-4-3
Complia
nce level
3 V
rms
(for main
power
ISM
line)
1V
rms
(for
probe)
3 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the CMS70A, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
é
ù
3
5 .
=
d
P
ê
ú
V
ë
û
1
é
ù
3
5 .
80 MHz to 800 MHz
=
d
P
ê
ú
E
ë
û
1
é
ù
7
800 MHz to 2.5 GHz
=
d
P
ê
ú
E
ë
û
1
where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in
each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
25
b
a