Télécharger Imprimer la page

Miele MF01 10 Notice De Montage page 35

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 9
Produktgruppe 512
35 von 38
Danger!
The warnings and instructions on detergent and disinfectant containers must be followed closely. The operator is responsible for ensuring
that all containers are clearly marked with suitable warnings when detergents and disinfectants are transferred from large containers into
smaller ones for use at the installation site.
Note
The DOS-P2 dispenser pump, mat. no. 05573160, contains a dispenser pump and includes the parts required for connection to a
washing machine:
Dispenser pump
Suction probe
Hoses, clips
Connection parts for connection to the machine
Installation instructions - Pump hose
Installation instructions - Liquid dispensing
Note
The DOS-R non-return valve, mat. no. 05571620, is required with the use of acidic agents, bleaches or similar chemicals, in order to
prevent backflow of chemicals (e.g., when containers are changed) into the suction pipework. It includes the following:
Non-return valve
Installation instructions - Non-return valve
Note
The DOS-G2 dispenser pump housing, mat. no. 05572530, is required for installing dispenser pump DOS-P2. The dispenser pump
housing is for maximum 3 dispenser pumps. It includes the following:
Housing with lid
Suction connections
Wiring material, fastenings
Wiring diagram - Liquid dispensing, WS 5101 - WS 5323
Wiring diagram - Liquid dispensing, PW 6101 - PW 6323
Wiring diagram - Liquid dispensing, PW 6107 - PW 6207
Note
The DOS-S dispenser stand, mat. no. 04814240, optional, for holding the DOS-G2 dispenser pump housing for maximum 3 dispenser
pumps includes the following:
Fastenings for the DOS-G2 dispenser pump housing and the spray proofing pump
List of illustrations:
Fig. 1, Drilling out/Shortening connection stubs
Fig. 2, Non-return valve
Fig. 3, Dispenser pump housing, complete
Fig. 4, Dispenser pump installation
Fig. 5, Dispenser pump stand
Disconnect the machine from the electrical supply provided in accordance with all relevant regulations and ensure that utilities
cannot be switched on again in error. Carry out a check to ensure that no power is supplied to the machine.
Remove the screws on the left and right of the top of the lid.
Lift the front of the lid slightly, slide it to the rear and thus release the centering cones at the rear of the lid from the connection strip.
Remove the lid upwards.
Cut and drill out (4 mm drill bit) the detergent dispenser box connection stubs as necessary according to the number of dispenser
pumps and the inner diameter of their connection hoses.
Warning!
Use the bottom connection stubs first.
Install the hoses on the connection stubs (Fig. 1, Pos. 1) and install hose clips (Fig. 1, Pos. 2) to secure them.
09.08.2011