18 Storage, cleaning, and maintenance
General compliance and standards
This product and its accessories must be disposed of according to local laws and
regulations. Do not dispose of this product as unsorted municipal waste. Prepare this
product for reuse or separate collection as specified by Directive 2002/96/EC of the
European Parliament and the Council of the European Union on Waste Electronic and
Electrical Equipment (WEEE). If this product is contaminated, this directive does not
apply.
For more specific disposal or compliance information, see www.welchallyn.com/weee,
or contact Welch Allyn Customer Service at +44 207 365 6780.
Troubleshooting
This section presents tables of information to help you troubleshoot issues with the
scale.
To use these tables, locate the specific problem with your scale in the left column of the
table. The remainder of the row explains possible causes and suggests actions that can
resolve the issue.
Issue
The display does not illuminate
The weight reading takes too long
to display
LOW BATTERY
The
indicator is illuminated
The scale displays
The scale displays
The scale displays
Directive 2002/96/EC-WEEE:
Disposal of noncontaminated electrical and electronic equipment
Possible cause
No power is supplied to the scale
The cable between the display board and the
instrument board is disconnected
The membrane keyboard and the display board
are disconnected
The cradle is in motion
The scale batteries are low on power
bAttrY
The batteries do not have enough power to
operate the scale
CAbLE
A signal outside the range of the internal A/D
converter has been applied
O-LOAd
The object on the scale is larger than the
maximum value allowed
Welch Allyn Pediatric/Infant Scale
Suggested action
Make sure that the scale is
plugged in.
Make sure that the batteries are
properly installed and are in good
condition.
Contact your service department.
Contact your service department.
Wait for the cradle to settle.
Replace the scale batteries.
Replace the scale batteries.
Contact your service department.
If the weight value is within the
specified range, contact your
service department.