3. RESISTANCE MEASUREMENT

1) Connect the 2 cables to the 4 measurement terminals, then the 2 Kelvin clips to the object to be
tested, which must not be live.
2) Set the switch to 400W - 10 mA.
3) Start the measurement
by pressing the START/
STOP button.
SET-UP
400
10mA
40
10mA
4000m
100mA
400m
1A
40m
1A
4000µ Ω
10A
OFF
Charge
4) Press the START/STOP
button again to stop the
measurement...
SET-UP
400
10mA
400 Ω
10mA
+I
mA
2nd
START/STOP
START/STOP
9
C
1
P
1
P
2
C
2
+I
If the measurement is too low, turn
the switch to the next lower range
and restart the measurement.
Continue until the display shows
at least 3 digits.
+I
2nd
+I
HOLD
2nd
A
A
mΩ
A
mΩ