The User Menu allows the customizing of balance
settings.
NOTE: Additional sub-menus may be available
if Interface Options are installed. See Interface
User Manual for the additional setting
information.
MENU:
C.A.L
Zero2
1
Span1
Span2
1
Lin1
Lin2
1
GEO
Menu
End
Items:
Only visible when dual scale is set to On. Some units
1
may not be available in all models.
MENU:
P.r.i.n.t.1
Reset
Stable
A.Print
Menu
Contnt
Items:
Layout
Data.Tr
End
NOTE: When Legal For Trade is set to on
(LFT switch in locked position), the menu
settings are affected as follows:
• Calibration (C.A.L) menu is not accessible.
• Dual Scale, Capacity 2, Graduation 2, Filter
and Units are locked at their current settings.
Zero Range setting is locked at 2%.
• Stable Range setting is locked at 1d.
• Auto-Zero Tracking setting is locked at 0.5d.
• Stable Only is locked On.
• Auto Print/Continuous is disabled.
• Lb:oz is locked Off.
PAGE 8 OF 69

MENU SETTINGS

S.E.t.U.P
r.E.A.d
Reset
Reset
D.Scale
Stable
Cap2
Filter
1
Grad2
AZT
1
Pwr. Un
Light
Zero
A.Off
A.Opt
End
A.Tare
Bp.Opt
Bp.Sig
Bp.Key
Accum
End
C.O.M.1
L.O.C.k
Reset
L.Setup
Baud
L.Read
Parity
L.Unit
Stop
L.Print
Handsh
L.COM
Alt.Cm
End
End
Summary of button navigation functions in menu mode:
• Yes: Allows entry into the displayed menu.
Accepts the displayed setting and advances to
o
the next item.
• No: Skips by the displayed menu.
Rejects the displayed setting or menu item and
U.n.i.t
o
advances to the next available item.
Reset
kg
• Back: Moves backward through the upper and
g
middle level menus.
lb
oz
Backs out of a list of selectable items to the
lb:oz
o
previous middle level menu.
End
• Exit: Exits from menu directly to the active weighing
mode.

CALIBRATION MENU

Enter this menu to perform calibrations.
When the scale is operated for the first time, a span
calibration is recommended to ensure accurate
E.n.d
weighing results. Before performing the calibration, be
sure to have the appropriate calibration weights as
listed in table. Ensure that the LFT switch/calibration lock
is set to unlocked position (See Figure 6 on page 16) or
adjust the GEO setting according to user's location.
PROCEDURE
Press and hold Menu until [mMeNU] is displayed. When
the button is released, the display will show [C.A.L]. Press
Yes to accept. [SpaN] will then be shown. Press Yes to
begin the span calibration. [ 0 kg] will be displayed.
Press Yes to accept. [--C--] will be displayed while zero
reading is stored. Next, the display shows the calibration
weight value. Place the specified calibration mass on
the pan. Press Yes to accept the weight or No to select
an alternate weight. [--C--] will be displayed while the
reading is stored. The display will show [done] if the
calibration was successful. The scale returns to the
previous application mode and is ready for use.
REQUIRED SPAN CALIBRATION MASS (SOLD SEPARATELY)
CAPACITY
1500g
3000g
6000g
NOTE: When active unit is g or kg, the
calibrating unit will be in kg. When active unit
is lb, oz or lb:oz, the calibrating unit will be in lb.
For linearity calibration, the calibration mass is
fixed. Mid-point is always half of full capacity.
MASS
CAPACITY
1.5kg / 3lb
15000g
3kg / 6lb
30000g
6kg / 15lb
MASS
15kg / 30lb
30kg / 60lb
1021 IH-9554