Probe setup
Fig 30
38
Fig 31
Fig 32
HPS50 PersonalScope
Probe
PROBE SETUP
x1 / x10
Press the 'Probe x1/x10'-key to set the measurements
accordingly the x1/x10-probe setting. (Fig. 30)
Notes:
• Automatically calculate the correct readouts
depending the x1 or x10 probe setting.
• An 'x10' symbol is displayed if this mode has
been selected.
• X10 measuring probes should be calibrated!
• IMPORTANT: Set the measuring probe in
the x10 position for measuring high voltages
(>100Vp+dc)
Setting up a X10 measuring probe :
When used in the X10 position, a measuring probe
should always be calibrated to the measuring ins-
trument being used, in this case the Personal
Scope. (Fig. 31)
Set the scope to X10 position
(probe x1/x10 key).
Set the voltage per division to 1V.
Set the time per division to 0,1ms.
Select AC for the input.
Use the probe to perform measurements at the
preselected point. Adjust the trimmer of the measu-
ring probe in order to obtain a square wave signal
with a top that is as flat as possible (Fig. 32)