Outline and Dimensions
Ferrule Considerations
The primary customer requirement is
a 4", 3A compliant sanitary ferrule
welded to a short standpipe on the
top of the tank. The inner standpipe
wall and ends must be smooth and
free of burrs, ridges, or seams, and
must comply with all applicable 3A
regulations.
The standpipe extension between
the mating ferrule and the tank top
cannot be longer than the blanking
distance as determined by the
Siemens Milltronics transceiver. The
blanking distance is the minimum
possible measurement range listed in
the Specifications on page 4.
Do not over tighten the mounting.
Hand tightening of the mounting
hardware is sufficient.
7ML19985HX61
XCT-8 Sanitary Transducer– INSTRUCTION MANUAL
1" BSP
or
1" NPT
119 mm
(4.68")
122 mm
(4.8")
radiating PTFE face
XCT-8
transducer
sanitary
clamp
3A compliant
standpipe
topped with 4"
ferrule
blanking
distance
Page EN-5