Declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of device should assure that it is used in such an environment.
Immunity
IEC 60601
Compliance level
test
test level
3V
0.15 MHz to
80MHz
Conducted
6 V in ISM
RF
and amateur
Not applicable
IEC 61000-
radio bands
4-6
between 0.15
MHz and 80
MHz
10V/m
Radiated RF
80 MHz to 2.7
IEC 61000-
10V/m
GHz
4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
42
Visiomed - KINECARE - 10-2017
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of
device, than the recommended
separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
[
[
3.5
d =
p
150 KHz to 80 MHz
V
1
[
[
3.5
d =
p
E
80 MHz to 800 MHz
1
[
[
3.5
d =
p
80 MHz to 2.7 GHz
E
1
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey, a
should be less than the compliance
level in each frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Field strengths from fixed RF transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which device is used exceeds
the applicable RF compliance level above, device should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating device.
Over the frequency range 0.15 MHz to 80 MHz, field strengths should be less than 3 V/m.
Table 4
Recommended separation distances between
portable and mobile RF communications equipment and device
Device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and device, as recommended below,
according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum
output power of
transmitter
0.15 MHz to 80 MHz
W
d = 1.2 p
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
m
80 MHz to 800 MHz
80 MHz to 2.7 GHz
d = 1.2 p
d = 2.3 p
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23
Visiomed - KINECARE - 10-2017