EMC data
Important information regarding electromagnetic compatibility (emc).
Table 1
Declaration - electromagnetic emission
The device is intended for use in the electromagnetic environment specified below. The
customer or the user should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR11
Harmonic emissions
Not applicable
lEC 61000-3-2
Voltage fluctuations /
flicker emissions
Not applicable
lEC 61000-3-3
Table 2
Declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user should assure that it is used in such an environment.
IEC 60601
Immunity test
test level
± 8 kV contact
Electrostatic
± 2 kV, ± 4 kV,
discharge (ESD)
± 8 kV,
lEC 61000-4-2
± 15 kV air
40
Visiomed - KINECARE - 10-2017
Electromagnetic environment -
guidance
The device uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
Complies
The device is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Electromagnetic
Compliance level
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
Compliance
synthetic material, the
relative humidity should
be at least 30 %.
±2kV for power
Electrical fast
supply
transient/burst
lines
IEC 61000-4-4
±1kV for input/
output lines
± 0.5kV, ± 1 kV
line(s) to lines
Surge
± 0.5kV, ± 1 kV,
IEC 61000-4-5
± 2 kV line(s) to
earth
0 % UT; 0.5
cycle At 0°, 45°,
90°, 135°, 180°,
225°, 270°and
315°
Voltage dips, short
interruptions and
0 % UT; 1 cycle
voltage
and
variations on power
70 % UT; 25/30
supply input lines
cycles
IEC 61000-4-11
Single phase:
at 0°
0 % UT;
250/300 cycles
Power frequency
(50/60Hz) magnetic
30 A/m
field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Not applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.
Not applicable
Mains power quality
should be that of a typical
Not applicable
commercial or hospital
environment.
Power frequency
magnetic fields should
be at levels characteristic
Complies
of a typical location in
a typical commercial or
hospital environment.
Visiomed - KINECARE - 10-2017