7. TROUBLESHOOTING GUIDE

Problem
Possible cause
The device does not switch on
The batteries are discharged
The batteries are not properly
installed
Treatment appears to be different
The patch is positioned incorrectly
or less pleasant than previously
(wrong place, wrongly applied, etc.)
The intensity is too low or too high
The electrodes' expiry date is
exceeded
The patch is deteriorated or worn
The patch is dirty
Unpleasant sensation when
The session was not stopped
touching the electrode or device
correctly (the on/off button was
fasteners
not pressed before removing the
device).
Your skin becomes red and/or you
The patches are always used in the
feel pain
same place
The patch does not adhere properly
to your skin
The patch is dirty
The surface of the patch is damaged
24
Visiomed - KINECARE - 10-2017

8. TECHNICAL FEATURES

Solution
Brand
Put new batteries in the
Model
device
Units
Install the batteries
Programs
properly, following the
correct polarity
Classification
Switch off the device
Intensity levels
and reposition the patch
Power supply
on your skin
Output voltage
Change the intensity
using the + or - buttons
Operating conditions
Replace the patch
Storage and transport conditions
Replace the patch
Electrostimulator dimensions
Clean the patch ,if the
problem continues,
replace the patch
Weight
Always switch off the
Lifespan of the batteries
device at the end of the
session before removing
Service life of device
the electrodes.
Press the STOP button
IP rating
for 2 seconds.
Waveform
Position the patches
differently
Pulse width
Make sure the patches
Frequency
are positioned correctly
Treatment time
on your skin
Electrode dimensions
Clean the patch If the
problem continues,
replace the patch
Replace the patch
KINECARE®
VM-WT1 Tens/EMS (wireless)
1 device and 1 patch (reference VM-MP1)
5 programs (3 TENS, 1 EMS, 1 Massage)
Medical device Class IIa
0 to 15
2 AAA batteries
45V (charged) at 500 ohm
5°C - 40°C / 41°F - 104°F with a relative humidity of 30% to
75% and atmospheric pressure of 700 to 1060 hPa.
-10°C - 55°C / 14°F - 131°F with a relative humidity of 10% to
90% and atmospheric pressure of 700 to 1060 hPa.
71mm(L)x55mm(w) x18.5mm(H)
2.79in(L)x2.16in(l)x0.72(H)
38g / 0,08lb (without batteries)
Approx. 30 uses (if used 30 min/day at half the maximum
intensity with new batteries on program 1)
2 years
IP22
Biphasic square wave
30 µs-250 µs
2Hz-290 Hz
30 minutes
210mm(L)x68mm(w)
8.26in(L)x2.67in(l)
Visiomed - KINECARE - 10-2017