Guidance And Manufacturer's Declaration - Electromagnetic Emissions; Guidance And Manufacturer's Declaration - Electromagnetic Immunity - Freestyle LibreLink Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

ART41361-001_rev-A 2:18 pm 30 aug 2019
interference to radio communications. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to radio or television reception,
which can be determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
— Reorient or relocate the receiving antenna.
— Increase the separation between the equipment and receiver.
— Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
— Consult the dealer or an experienced radio/TV technician for help.
Under Industry Canada regulations, this radio transmitter may only operate using an antenna of a type
and maximum (or lesser) gain approved for the transmitter by Industry Canada. To reduce potential radio
interference to other users, the antenna type and its gain should be so chosen that the equivalent
isotropically radiated power (e.i.r.p.) is not more than that necessary for successful communication.
This device complies with Industry Canada's licence-exempt RSS standard(s). Operation is subject to the
following two conditions:
(1) This device may not cause interference; and
(2) This device must accept any interference, including interference that may cause undesired operation of
the device.
Guidance and manufacturer's declaration – electromagnetic emissions
The Sensor is intended for use in the electromagnetic environment specified below. The customer or the
user of the Sensor should assure that it is used in such an environment.
Emissions test: RF emissions; CISPR 11
Compliance: Group 1
Electromagnetic environment – guidance: The Sensor uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
Emissions test: RF emissions; CISPR 11
Compliance: Class B
Electromagnetic environment – guidance: The Sensor is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low voltage power supply network
that supplies buildings used for domestic purposes.
Guidance and manufacturer's declaration – electromagnetic immunity
The Sensor is intended for use in the electromagnetic environment specified below. The customer or the
user of the Sensor should assure that it is used in such an environment.
Immunity test: Electrostatic discharge (ESD); IEC 61000-4-2
IEC 60601 test level: ± 8 kV contact; ± 15 kV air
Compliance level: ± 8 kV contact; ± 15 kV air
Electromagnetic environment – guidance: Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at least 30%.