Meter Check Disk
Meter Check Disk in chamber
instead of reagent disk
"Source" appears on
Wrong Market Type was chosen
test screen instead of
"Sanitizer"
Problem with default blank due
!
On Test Results and
to under filled disk or air bubble.
Test History screens
Test results are questionable.
Range Error
Raw data out of range
Output Error
Decreased light intensity.
Possibly dirty lens
Consistently unexpected
Metals may actually be present
high results for metals
Alkalinity result of 0 ppm
Usually due to an underfilled
disk.
Unexpected results
Dirty disk cover
Disk type is not an option
Software or meter firmware is
in Disk Series
out of date.
High pH results
Water sample temperature
above 100 °F (38 °C) interferes
with pH reagent
Salt results not as
Water sample not between
expected
68-74 °F (20-23 °C). Sample
disk temperatures above or
below this will also cause
artificially low or high readings
respectively.
Disk not spinning
Lid open, meter not powered
on, low battery, disk or disk
cover pressed down too tightly
on hub
Fast electrical transients may
disrupt operation of the Spin
Touch meter
Test result value on the
Results are out of range of
display is red
reagent system
Trouble connecting to
Bluetooth not enabled
Bluetooth device
Too many Bluetooth devices
near the meter
Printer power is ON and Fast
Printer Connection is ON
GlobalTestSupply
www.
Select "Continue" to go to Test
Results screen. Select "Abort" to go
to Testing screen and run test with
reagent disk.
Go to Settings. Change Market Type
to Pool/Spa
Fill disk correctly (see FILLING). On
Test Results screen tap ! for details.
Contact Support
Clean lens (see CLEANING). Follow
Range Check Procedure. If error
message persists, contact Support.
Repeat test with distilled water. If
the results still show that metals are
present, contact Support.
Review the disk filling procedures
and test again. If problem persists,
contact Support.
Gently clean disk cover apertures
with pipe cleaner or lint free cloth.
For water samples over 100 °F
(38 °C) subtract 0.3 from pH result
or, for the most accurate result, wait
until water sample is below 90 °F
(32 °C) to test
Run salt test with sample water and
disk between 68-74 °F (20-23 °C).
Close lid, power on meter, charge the
battery or plug meter into a stable
power source, remove the disk/disk
cover and place back in the chamber
more gently
Restart the test to resume normal
operation
Dilute sample. Retest for out of
range test factor. (All factors except
pH.)
Enable Bluetooth
Have only one device near the meter
Turn printer power OFF. Or turn Past
Printer Connection OFF.