Télécharger Imprimer la page

Intended Use; Contraindication; Product Description; Specifications - Lanaform WBPM-110 Manuel D'instructions

Sphygmomanomètre
Masquer les pouces Voir aussi pour WBPM-110:

Publicité

Les langues disponibles

Les langues disponibles

EN
FR NL DE
1
Day and time display
2
Systolic
3
Diastolic
4
Pulse rate
5
Irregular heartbeat symbol
6
Low battery indicator
7
The pressure in cuff is instable,
or much remnant air in cuff
8
Blood pressure level classification indicator
9
Blood pressure level classification indicator
10
« MEM » button
11
« START button »
12
LCD
13
Cuff

3. INTENDED USE

Fully Automatic Electronic Sphygmomanometer is
for use by medical professionals or at home and is a
non-invasive blood pressure measurement system
intended to measure the diastolic and systolic blood
pressures and pulse rate of an adult individual by
using a non-invasive technique in which an inflatable
cuff is wrapped around the wrist.

4. CONTRAINDICATION

report_p It is inappropriate for people
with serious arrhythmia
to use this Electronic
Sphygmomanometer.

5. PRODUCT DESCRIPTION

Based on Oscillometric methodology and silicon inte-
grated pressure sensor, blood pressure and pulse rate
can be measured automatically and non-invasively.
The LCD display will show blood pressure and pulse
rate. The most recent 2×60 measurements can be
stored in the memory with date and time stamp. The
Electronic Sphygmomanometer corresponds to the
4
/ 64
WBPM-110
below standards: IEC 60601-1Edition 3.1 2012-08/EN
60601-1:2006/A1:2013 (Medical electrical equipment
-- Part 1: General requirements for basic safety and
essential performance), IEC60601-1 2:2014/EN 60601-
1-2:2007 /AC:2010 (Medical electrical equipment
-- Part 1-2: General requirements for basic safety and
essential performance - Collateral standard: Electro-
magnetic compatibility - Requirements and tests),
IEC80601-2-30:2009+AMD1:2013/EN 80601 2-30:2010/
A1:2015(Medical electrical equipment –Part 2-30:
Particular requirements for the basic safety and
essential performance of automated non-invasive
sphygmomanometers)EN 1060-1: 1995 + A2: 2009
(Non-invasive sphygmomanometers - Part 1: General
requirements), EN 1060-3: 1997 + A2: 2009 (Non-in-
vasive sphygmomanometers - Part 3: Supplementary
requirements for electro mechanical blood pressure
measuring systems);ISO81060-2 :2013(Non-Invasive
Sphygmomanometers - Part 2: Clinical Validation Of
Automated Measurement Type).

6. SPECIFICATIONS

1
Product name: WBPM-110
2
Model: LA090207
3
Classification: Internally powered,
Type BF applied part, IP22, No AP
or APG, Continuous operation
4
Machine size: Approx.83mm X 74mm
X26mm( 3 9/32''x 2 29/32''x 1 1/32'')
5
Cuff circumference: 14cm-
19.5cm (5 1/2''-7 11/16'')
6
Weight: Approx71g (2 1/2oz.)
(exclude batteries and cuff)
7
Measuring method: Oscillometric method,
automatic inflation and measurement
8
Memory volume: 2×60 times
with time and date stamp
9
Power source: batteries: 2×1.5V SIZE AAA
10
Measurement range:
a
Cuff pressure: 0-300mmHg
b
Systolic: 60-260mmHg

Publicité

loading

Ce manuel est également adapté pour:

La090207