Average Gas Meter Γ Factor - PFS 5520 Mode D'emploi

Table des Matières

Publicité

Meter Manufacturer:
Model:
Lab ID #:
Serial #:
Calibration Date:
Calibration Expiration:
Barometric Pressure: 29.97
Reference Standard DGM
Manufacturer: Apex
Model: SK25DA
Lab ID#: 47
Serial #: 1101001
Calibration Expiration Date: 3/13/2020
Calibration γ Factor: 0.998
Calibration Data
Standard DGM Initial Volume (L)
Standard DGM Final Volume (L)
Standard DGM Temperature (°F)
Standard DGM Pressure (in H
3
DGM Initial Volume (ft
)
3
DGM Final Volume (ft
)
DGM Temperature (°F)
DGM Pressure (in H
O)
2
Time (min)
Net Volume for Standard DGM (ft
3
Net Volume for DGM (ft
Dry Gas Meter γ Factor
γ Factor Deviation From Average
Average Gas Meter γ Factor
Calculations:
1. Deviation = |Average value for all runs - current run value|
2. γ = [V
x (γ
) x (P
std
Std
bar
Standard Reference Meter is calibrated to NIST traceable standards. Uncertainty of measurement is ±0.5%.
PFS-TECO
Dry Gas Meter Calibration
Apex
XC-60-ED
53
1902130
6/14/2019
12/14/2019
O)
2
3
)
)
+P
/13.6) x (T
+ 460)] / [V
std
DGM
in. Hg
Unit Under Test Previous Calibration
Allowable Deviation (±5%):
Run 1
0.000
147.373
71.0
0.00
0.000
5.227
78.0
2.67
33.0
5.204
5.227
1.000
1.000
0.999
x (T
+ 460) x (P
DGM
std
Page 1 of 1
Date
12/17/2018
γ Factor:
1.004
0.0502
Actual Deviation:
Result:
PASS
Run 2
Run 3
0.000
0.000
142.005
143.359
72.0
0.00
0.000
0.000
5.114
5.224
85.0
2.00
36.0
5.015
5.063
5.114
5.224
0.998
0.998
0.998
0.998
+ P
/ 13.6)]
bar
DGM
Technician:____________________
0.01
72.0
0.0
91.0
1.5
42.0