Philips Avent SCF334 Mode D'emploi page 39

Masquer les pouces Voir aussi pour Avent SCF334:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Immunity
IEC
test
60601
test
level
Conducted
3 Vrms
RF IEC
150 kHz
61000-4-6
to 80
MHz
Radiated RF
3 V/m
IEC 61000-
80 MHz
4-3
to 2,5
Ghz
4213_354_3951_4_DFU-Luxury_schalen naar 140x200_v7.indd 39
Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic
environment specified below. The user of the Philips Avent breast pump
should assure that it is used in such an environment.
Com­
Electromagnetic environment – guidance
pli ance
level
3 Vrms
Portable and mobile RF communications equipment should be
used no closer to any part of the Philips Avent breast pump,
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance d=1,17 √P The breast pump
can be turned off.
80 MHz to 800 Mhz d=1,17 √P
3 V/m
800 MHz to 2,5 GHz d=2,33 √P
where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey, should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with the following symbol:
Caution: Although the breast pump is compliant with applicable
EMC directives, it may still be susceptible to excessive
emissions and/or may interfere with other equipment. A
consequence can be that the breast pump turns off or the
on/off button starts flashing (see troubleshooting for more
information), which is acceptable as this will not lead to
unacceptable risks. To prevent interference, keep other electric
equipment away from the breast pump during expressing.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Philips Avent breast
pump is used exceeds the applicable RF compliance level above, the Philips
Avent breast pump should be observed to verify normal operation.
English
39
26/07/2018 15:20

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Avent scf332Avent scd292Avent scd223/20Avent scd223/00

Table des Matières