Problem
Causes
Coupler or endoscope not
properly focused
Endoscope not seated on
coupler
Image out of focus
Debris on endoscope tip
Debris on device window or
coupler optics
Moisture/debris on endoscope
tip
Foggy Image
Moisture/debris on device
window or coupler optics
Enhance level too high
Grainy image
Firmware fault
Device not connected to
Camera Control System
Buttons on device
Firmware fault
do not function
Buttons damaged
Improperly connected keyboard
Invalid keyboard
Keyboard does
not function
Keyboard not functioning
Firmware fault
Improperly connected USB
flash drive
Images or video
Invalid USB flash drive
not storing on
USB flash drive
Firmware fault
USB flash drive full
Possible Solution
Focus coupler following coupler IFU
instructions
Remove endoscope from coupler and
re-attach
Clean endoscope tip
Remove coupler from device and clean
glass surfaces on device and coupler
Clean endoscope tip
Remove coupler from device and clean
glass surfaces on device and coupler
Reduce Enhance Level
Reset Camera Control System
Connect device to Camera Control
System
Reset Camera Control System
Consult Warranty/Service Return
information
Connect keyboard correctly
Connect correct keyboard
Repair or replace keyboard
Reset Camera Control System
Connect USB flash drive correctly
Connect a USB Flash Drive FAT 32
format
Reset Camera Control System
Remove images from USB flash drive
using a computer and retry .