Les langues disponibles

Pulse Generator
Input (PPM)
31,423
59,335
127,043
177,752
205,031
5. Turn off the Inspector+. Hold down the "-" button on the end panel and turn
on the Inspector+. The display shows CAL, and the Inspector+ counts down
for 15 seconds, chirping each second. At the end of the 15 seconds, the
Inspector+ beeps.
6. The Inspector+ collects data for 30 seconds, chirping every 2 seconds, with
CAL and the hourglass icon flashing. At the end of the 30 seconds, it beeps.
The display shows CAL, and SET is flashing
7. Press the "+" and "-" buttons to adjust the reading to what it should be.
When the reading is correct, press the Set button.
The new calibration factor is displayed for several seconds, then the
Inspector+ beeps and resumes normal operation.
Troubleshooting
The Inspector+ is a highly reliable instrument. If it does not seem to be working
properly, look through the following chart to see if you can identify the problem.
Problem
Display is blank
20
CPM
mR/Hr
33,400
10.00
66,800
20.00
166,999
50.00
267,200
80.00
334,000
100.0
Possible Cause
no battery, dead battery,
poor battery connection
defective LCD
µ µ Sv/hr
CPS
100.0
557
200.0
1,113
500.0
2,783
800.0
4,453
1,000
5,567
What To Check
install a new 9-volt battery
if count light and audio
work, the LCD may
need to be replaced