Instructions For Use

Preparation and Setup of the Ranger Irrigation Warming Unit

1. Attach the Ranger warming unit to the I.V. pole. Tighten the pole clamp
securely.
:
To prevent tipping, clamp the Ranger Irrigation warming unit
caution
to an I.V. pole with a minimum 14" (35.6 cm) radius wheelbase and at a
height no higher than 44" (112 cm). Failure to do so may result in damage
to the product (see figure 1).
2. Slide the irrigation fluid warming cassette into the slot in the warming
unit. The cassette can only fit into the device one way (see figure 2).
3. Prime the disposable set. For more information about priming the set,
refer to instructions provided with the disposable sets.
4. Plug the power cord into an appropriate outlet. Turn the unit ON. In a
few seconds the alphanumeric display will illuminate. It takes less than 2
minutes to warm up to the 41
5. Begin infusion. When infusion is completed, remove the disposable set
and discard according to institutional protocol.
Removing the Irrigation Fluid Warming Disposable Set
from the Ranger Irrigation Warming Unit
1. Close the inlet clamp proximal to the fluid warming cassette and open all
clamps distal to the cassette.
2. Disconnect the disposable set from the fluid source, if applicable.
3. Allow a small amount of fluid to drain from the end of the patient line to
reduce pressure in the cassette.
4. Remove the irrigation fluid warming cassette from the warming unit and
discard according to institutional protocol.
Ranger Irrigation Fluid Warming System
C set point temperature.
°
English
Figure 1
Figure 1
44"
44"
14" (35.6 cm)
14" (35.6 cm)
(112 cm)
(112 cm)
Figure 2
Figure 2
7